Alarms
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
BATTERY MISSING
Displays when the Pump cannot detect an installed Battery Module.
■
Confirm that the Battery Module is properly installed and securely latched into
the Pump.
CLEAN LOAD POINT
#2
Displays when the Pump detects foreign matter in the tubing channel at Load Point 2.
1.
Observe there is no debris or foreign matter at Load Point 2.
2.
If observed, remove debris and clean the tubing channel. See “Cleaning the
Pump” on page 118.
3.
If the alarm repeats, return the Pump for service to qualified personnel at your
facility.
CLOCK BATTERY
LOW
Displays when the real-time (internal coin cell battery) clock (RTC) battery is low.
■
Return the Pump to qualified personnel at your facility or to Baxter for service.
CLOSE CLAMP
RELOAD SET
Displays when an IV set is loaded into the Pump before the Pump has completely
powered on.
1.
Unload the IV set.
2.
Press
ON/OFF
to power off the Pump.
3.
Press
ON/OFF
to power on the Pump.
4.
After the Pump has completely powered on, reload the IV set. See “Loading an
IV Set” on page 37.
DOOR NOT FULLY
LATCHED
Displays when the Pump’s door is closed but not fully in the locked position.
1.
Close the roller clamp below the Pump.
2.
Open the door by inserting the slide clamp into the keyhole.
3.
Re-load the IV set following the on-screen prompts.
4.
Close the Pump’s door by pressing the upper and lower corners near the door
latches. See “Loading an IV Set” on page 37.
DOOR OPEN
Displays when the Pump’s door is open while an IV tube is loaded in the tubing
channel.
1.
Close the roller clamp.
2.
Assess the IV line for air.
3.
Close the door.
OR
■
Unload the
tubing by pulling it out and up from the bottom of the Pump
towards the top. See “Unloading an IV Set” on page 44.
Alarms