Programming the Pump
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
1. To stop the delay timer, press
STOP
. The display updates to
DELAY STOPPED
and the delay
timer is paused and no longer flashes. To restart the delay timer, press
RUN
.
2. To cancel the delay, press the
cancel
soft key. The remaining delay time is cleared and the
display updates to
PUMP STOPPED
.
To change the remaining delay time:
1. Press
STOP
.
2. Press the
review/program
soft key to display the setup screen.
3. Move the cursor to the delay value, enter the new desired delay time and press
OK
.
NOTE:
The new delay time is immediately applied. The delay time value may not be
cleared while the delay is running.
4. Press
RUN
to start the delay timer.
Confirming Weight and BSA
The Weight
or BSA confirmation screen (set up in the Drug Library), requires the patient weight or
BSA parameter to be re-entered to confirm that the value was entered correctly. If the weight or
BSA values do not match, the Values Differ popup message is displayed. Press
OK
and enter an
accurate value. See Figure 80. Also see Figure 81.
NOTE:
Important
. For the drug products mentioned, refer to the manufacturers’ package insert
for full prescribing information which may include boxed warnings.
The flow rate (mL/hr) will be auto calculated based on the patient weight and dose entered.
Amount/Time Infusion
Amount/Time programming is used for drug therapies that require a prescribed drug dose amount
to be delivered over a set duration of time.
Figure 80. Weight Confirmation Screen.
Figure 81. BSA Confirmation Screen. See
important note on page 80.