Programming the Pump
83
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
To display infusion status while the Pump is infusing, press the
review/edit VTBI
soft key
from the Infusion Running screen. See Figure 84. Also see Figure 85.
To change the mL/hr rate while stopped, press the
review
or
program pri/sec
soft key.
Press the arrow soft key to scroll up to change the rate.
To change mL/hr rate while running, press the
rate change
soft key and enter the new
rate.
To program a Standard Concentration - Drug Amount/mL:
The standard concentration may not be changed at the Pump. For example: 4 mg/mL.
1. Enter drug amount. Press
OK
.
2. Confirm standard concentration and press
OK
. See Figure 86. Also see Figure 87.
NOTE:
VTBI is automatically calculated and displayed.
NOTE:
After
RUN
is pressed, the VTBI cannot be edited in any Amount/Time infusion.
3. Enter duration of infusion. Press
OK
. See Figure 88.
Dose is calculated and displayed. See Figure 89.
Figure 86. Standard Concentration
Setup Screen.
Figure 87. Standard Concentration
Confirmation Screen.