Appendix E: Electromagnetic Compatibility
151
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
APPENDIX E: ELECTROMAGNETIC
COMPATIBILITY
Emissions
WARNING
Emissions and Immunity.
The use of accessories or cables other than those specified by Baxter
may result in increased Emissions or decreased Immunity of this
medical device.
WARNING
Radio Frequency Interference.
The SIGMA Spectrum Infusion System meets the electromagnetic
compatibility (EMC) requirements as specified in the International
Electrotechnical Commission’s (IEC) 60601-1-2 (2001-09) standard
for emissions and immunity. There may be potential difficulties if the
Pump is not kept separated from other equipment, such as hand-held
transmitters, cellular phones and electrosurgical equipment that may
generate strong radio frequency interference (RFI). See “Immunity –
Separation Distances” on page 156 for the recommended minimum
distance.
Guidance and manufacturer’s declaration – electromagnetic emissions –
for all equipment and systems
The SIGMA Spectrum Infusion System is intended for use in the electromagnetic environment specified below. The
customer or user of the Pump should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The SIGMA Spectrum Infusion System uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The SIGMA Spectrum Infusion System is suitable for use in
all establishments other than domestic and those directly
connected to a low-voltage power supply network that
supplies buildings used for domestic purposes.