Programming the Pump
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
3. Enter the desired flush amount to be delivered and press
OK
The Pump will deliver the Line Flush at the mL/hr rate of the previous infusion.
NOTE:
If additional action is required, to change the Line Flush amount (mL) after entering and
confirming it, press the
edit
flush
soft key.
4. To start the infusion, ensure that the slide clamp and roller clamp are open and press
RUN
.
Once
RUN
is pressed, the infusion will start and the Secondary Check Flow screen is
displayed. Check the flow and confirm:
All clamps are open.
There are no kinks or collapses in the tubing outside of the Pump.
Drops are flowing in the drip chamber. Note, at very low rates, it may take several minutes
to see drops.
Vents are open (if applicable).
If drops are falling in the drip chamber, press the
yes
soft key.
If no drops are falling, press the
no
soft key and follow the screen prompts.
Figure 110. Enter Secondary Line Flush
Volume. See important note on page 91.