Programming the Pump
65
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
■
If
REQUIRED
:
The Pump will display the Callback dialog box and the Pump will drop to the KVO Rate at the
completion of the secondary infusion and sound an audio alarm. This alarm must be
acknowledged before making a transition to the primary infusion. Review the primary infusion
or return to and manage the secondary setup.
■
If
OPTIONAL
:
The Pump will display the Callback dialog box upon completion of programming the secondary
infusion. If during the infusion, secondary callback is desired, press the
info settings
soft key
to change callback status during secondary infusion. Press the
yes
soft key to receive a
callback at the end of the infusion. Press the
no
soft key to automatically transition to the
primary infusion.
■
If
NEVER
:
The Pump will automatically transition to the primary infusion upon completion of the
secondary infusion.
To select Secondary Callback:
1. When finished programming a secondary infusion, if Secondary Callback is enabled as
optional (per drug in the Drug Library), a dialog box is displayed asking if a callback is needed
at the completion of the secondary infusion. See Figure 57.
If
yes
is pressed, a callback icon (a picture of a telephone) is displayed on the Secondary
Infusion Running screen to indicate that the callback feature is active. The Callback icon
alternates with the Secondary icon (picture of two IV bags) during the secondary infusion.
See Figure 58.