Appendix A: Specifications
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
Conversion Factors
Conversion factors are applied as appropriate to calculate rate or
dose (that is, 60 minutes = 1 hour, 1000 mcg = 1 mg, and so forth).
To calculate rate from an entered dose, the following formulas are
applied:
■
General:
■
Weight based:
To calculate dose from an entered rate, the following formulas are
applied:
■
General:
■
Weight based:
To calculate rate from an entered volume to be infused and entered
time of infusion, the following formula is applied:
External Interfaces
IrDA (SIR Encoding Protocol)
Flow Rate
0.5 to 999 mL/hr with 0.1 mL/hr increments from 0.5 to 99.9 mL/hr
and 1.0 mL/hr increments from 100 to 999 mL/hr
Infusion Delivery Modes
Continuous (Primary and Secondary), Multi-Step, Cyclic TPN, and
Amount/Time (Primary and Secondary)
KVO
At the completion of a primary infusion, the Pump will infuse at the
KVO rate configured per drug in the Drug Library or the current
infusion rate, whichever is lower. The default KVO rate is set at
1 mL/hr but may be configured to between 0.5 and 50 mL/hr.
At the completion of a secondary infusion program, the Pump will
run at a fixed KVO rate of 1 mL/hr.
Specification
Description
Rate
Dose
Concentration
--------------------------------------
=
Rate
Dose Patient Weight
Concentration
-----------------------------------------------------------
=
Dose
Rate Concentration
=
Dose
Rate Concentration
Patient Weight
---------------------------------------------------------
=
Rate
Volume to be Infused
Time of Infusion
---------------------------------------------------------
=