Compatible IV Sets
127
SIGMA Spectrum Infusion System
Pump Operating Software Version 8.00
Operator’s Manual
For Use With MDL Version 8.0
Buretrol, Clearlink, CONTINU-FLO and Interlink are all registered names/trademarks associated
with Baxter International Inc.
Compatible Baxter IV Sets – WARNINGS
NOTE:
The numbers in the warning correspond with the superscript reference numbers found
in the compatible IV set list.
WARNING
Baxter IV Sets.
1. Minidrip chambers should not be used for flow rate settings
greater than 200 mL/hr. Doing so may influence flow rate
accuracy and cause air in line or upstream occlusion alarms.
2. When using sets with backcheck valves, flow rate settings should
not exceed 500 mL/hr. Doing so may influence flow rate accuracy
or cause air in line or upstream occlusion alarms. Secondary flow
rates above 300 mL/hr may cause fluid to siphon from the
primary container.
NOTE:
Not applicable with non-DEHP tubing because 250 mL/hr is
the maximum flow rate per warning statement.
Buretrol Minidrip (60 drops/mL)
2H8862
CLEARLINK System, Non-DEHP BURETROL Solution Set, 115" (2.9 m), 150 mL
Valveless Burette, 4 Luer Activated Valves, Male Luer Lock Adaptor with Retractable
Collar
1, 5, 8, 11
2H8864
CLEARLINK System, Non-DEHP BURETROL Solution Set, 105" (2.7 m)150 mL Burette,
Drip Chamber Filter Valve, 3 Luer Activated Valves, Male Luer Lock Adapter with
Retractable Collar
1, 4, 5, 8, 11
IV Fat Emulsion Administration Set (10 or 60 drops/mL)
2C1145
Non-DEHP IV Fat Emulsion Administration Set, 10 drop/mL, 86" (2.2 m)
11
2C1146
Non-DEHP IV Fat Emulsion Administration Set, 60 drop/mL, 86" (2.2 m)
1, 11
Special Use (10 drops/mL)
2C8858
CLEARLINK System, Paclitaxel Set, 107" (2.7 m), Polyethylene Lined Tubing, Non-DEHP
Pump Segment, 0.22 Micron Downstream High Pressure Filter, Luer Activated Valve, Male
Luer Lock Adapter
3, 4, 7, 11
2C8875
CLEARLINK System, Solution Set with DUO-VENT Spike, 103" (2.6 m), Polyethylene
Lined Tubing, Non-DEHP Pump Segment, Luer Activated Valve, Male Luer Lock Adaptor
3,
7, 11
No.
Brief Description - Non-DEHP Sets