120
Cleaning and Storage
Manual 41018v0800
SIGMA Spectrum Infusion System
Revision B
Operator’s Manual
Cleaning the Battery Module
To clean the battery module:
1. Remove the Battery Module from the Pump.
2. Apply cleaning agent with a dampened cloth per the manufacturer’s instructions using
appropriate dilution ratio.
NOTE:
Do not spray solutions directly on the Battery Module or the exposed battery
terminals.
3. Carefully inspect and clean any debris from around the battery terminals.
4. Dry Battery Module thoroughly before replacing into the Pump.
5. Dispose of all cleaning materials as required per facility protocol/biohazard policy.
Cleaning the AC Power Adaptor
To clean the AC Power Adaptor:
1. Power off the Pump and unplug the AC Power Adaptor from the power source.
2. Disconnect the power cord from the Pump.
3. Apply cleaning agent to a dampened cloth per the manufacturer’s instructions using
appropriate dilution ratio. Do not spray solutions directly on the AC Power Adaptor.
4. Wrap the cloth on the AC Power Adaptor cord and clean the full length of the cord.
5. Wipe the power cord housing and the connector, and carefully clean the AC Power Adaptor
prongs.
6. Dry the AC Power Adaptor thoroughly.
7. Inspect the AC Power Adaptor for visual evidence of damage such as cracks, bent prongs,
cuts or exposed wires or dents:
AC Power Adaptor Module
AC Power Adaptor Cord
AC Power Cord Adaptor Retainer
8. If no signs of damage are found, plug the AC Power Adaptor into a working wall outlet.
If signs of damage are found, discontinue use and return the AC Power Adaptor to qualified
service personnel at your facility for service.
9. Verify green LED on the AC Power Adaptor is lit while plugged in and a plug icon appears on
the Pump’s display.
10. Dispose of all cleaning materials as required per facility protocol/biohazard policy.