5.6 Error correction and tips
Error
Error correction
No function
•
Check connection. Power supply 15…28 VDC on pin 2 (+Vs,
brown) and pin 7 (GND, blue)
Green LED flashes
•
Short circuit on switching outputs. Check connection
Red LED lights up
•
Object outside field of view (near, far, or to the side)
•
Amplitude of the received signal is insufficient (e.g., in case of
soiling)
Sensor does not provide the
expected measuring results
•
Check inclination angle and work in Flex Mount mode if required
(teach the new reference surface)
•
The object is not in the measuring range
•
Bright object, avoid direct reflexes from the transmitter to the
receiver
The sensor does not take
account of all objects within the
field of view
•
Enlarge field of view. The fie
ld of view was possibly limited; see
Section "FIELD OF VIEW"
•
The red visible laser beam does not represent the maximum field
of view. If the object is at the edge of this beam it could be outside
the measuring range
•
Move object. The object is outside the field of view vertically or is
in the blind region of the sensor
Unreliable measurement value:
The measurement value jumps
back and forth
•
The object is not in the measuring range
•
Avoid bright object
•
Avoid very dark object
•
Too much ambient light
•
Check measurement mode setting (MEASUREMENT MODE)
Transmitting laser light is dim
Sync-In input is on High--> set to Low
Incorrect measurement values
This is probably due to a malfunction caused by ambient light. Attempts
should be made to reduce the ambient light
A communication error occurs
during changing the IP address
If the IP address is changed, it must be reconnected to the sensor
No tool status message during
an error
A tool status message only states that the communication with the sensor
that was initiated by the call was successful. If there are problems, for
example when teaching with Flex Mount, this is returned by the sensor as
an error code