Using TheaterWatch
MGP D5 Dicom Theater Lite
97
Figure 39: Dual-head equalization
The Equalization option will be available only if:
•
the system contains at least one display of the same type as
the one being calibrated
•
at least one of the displays of the same type has already
been calibrated before using a sensor.
To equalize the displays:
1. Be sure the reference display is calibrated to the desired
luminance.
2. Set the Equalization option in the Application Settings dialog.
3. Close the Application Settings dialog.
4. Open the Display Settings dialog and calibrate display (2).
The display will be calibrated and additionally the display
luminance will be matched to the luminance of the reference
display (1).
BARCO
BARCO
c
d
(1) Reference display
(2) Calibrated display: Luminance automatically matched
to reference display
Summary of Contents for MGP D5
Page 1: ...Installation User Manual MGP D5 Dicom Theater Lite...
Page 3: ...Copyright notice 4 MGP D5 Dicom Theater Lite This page intentionally left blank...
Page 6: ...MGP D5 Dicom Theater Lite 7 Preface...
Page 14: ...MGP D5 Dicom Theater Lite 15 Overview...
Page 25: ...Quick installation overview This page intentionally left blank 26 MGP D5 Dicom Theater Lite...
Page 26: ...MGP D5 Dicom Theater Lite 27 Display Controller Installation...
Page 34: ...MGP D5 Dicom Theater Lite 35 Display Controller Software Installation...
Page 46: ...MGP D5 Dicom Theater Lite 47 Display Controller Tools...
Page 61: ...BarcoMed Hardware Tab 62 MGP D5 Dicom Theater Lite...
Page 62: ...MGP D5 Dicom Theater Lite 63 Display Tools...
Page 74: ...MGP D5 Dicom Theater Lite 75 MediCal Pro Software...
Page 78: ...MGP D5 Dicom Theater Lite 79 TheaterWatch Software...
Page 100: ...MGP D5 Dicom Theater Lite 101 Troubleshooting...
Page 110: ...MGP D5 Dicom Theater Lite 111 Warranty Statement...