Using TheaterWatch
94
MGP D5 Dicom Theater Lite
3. Click the
OK
button.
The calibration and the luminance will be reset to the latest
calibrated position.
Note:
When you start up the system, the calibration is also reset to
the latest calibrated position. The luminance, however, is
kept at the latest value.
To view a graph showing the result of the latest calibration:
1. In the Calibration tab dialog, click on
Show calibration
chart
.
This button is available only if the display has been
calibrated with a sensor before.
2. A graph is shown displaying the latest calibration compared
to the theoretical display function.
To revert to the default factory settings:
1. In the Calibration tab dialog, click on
Reset
.
2. As a result, the settings are restored to the default factory
values.
Summary of Contents for MGP D5
Page 1: ...Installation User Manual MGP D5 Dicom Theater Lite...
Page 3: ...Copyright notice 4 MGP D5 Dicom Theater Lite This page intentionally left blank...
Page 6: ...MGP D5 Dicom Theater Lite 7 Preface...
Page 14: ...MGP D5 Dicom Theater Lite 15 Overview...
Page 25: ...Quick installation overview This page intentionally left blank 26 MGP D5 Dicom Theater Lite...
Page 26: ...MGP D5 Dicom Theater Lite 27 Display Controller Installation...
Page 34: ...MGP D5 Dicom Theater Lite 35 Display Controller Software Installation...
Page 46: ...MGP D5 Dicom Theater Lite 47 Display Controller Tools...
Page 61: ...BarcoMed Hardware Tab 62 MGP D5 Dicom Theater Lite...
Page 62: ...MGP D5 Dicom Theater Lite 63 Display Tools...
Page 74: ...MGP D5 Dicom Theater Lite 75 MediCal Pro Software...
Page 78: ...MGP D5 Dicom Theater Lite 79 TheaterWatch Software...
Page 100: ...MGP D5 Dicom Theater Lite 101 Troubleshooting...
Page 110: ...MGP D5 Dicom Theater Lite 111 Warranty Statement...