Barco Display Tab
48
Display Properties
You may access the Properties page of the currently selected
display, which is represented by the monitor icon above with the
black frame around it, by either clicking on the
“Properties”
button or double clicking on the
icon
. This screen (figure 20 on
page 49) shows properties of the currently active display.
Name:
Displays the model name of the display.
Serial Number:
Displays the serial number of the display.
Backlight RunTime:
This is time in hours that the back light has
been on. A common question is: How much longer will the
backlight last? The backlight will typically last a very long time,
but will only be able to hold a calibrated output of for 17000
hours, after which time it will become slowly dimmer. For
displays calibrated at, an estimate of how long it will be before a
backlight replacement is needed = 17000 hours – backlight age.
Firmware Revisions:
This is the runtime firmware that is loaded
in the display. This information may be helpful when there is a
problem.
Internal Temperature:
This is the temperature inside the display.
It starts out at about the ambient temperature when the display
backlight is first turned on, then rises slowly to a steady-state
temperature as the display warms up. Changes in light level due
to temperature variations are completely controlled by the I-
Guard sensor on the front of the display.
Current Luminance Value:
This is the current luminance value of
the display. It may vary a bit from time to time, for the display
continuously calibrates itself to meet the target luminance value.
Summary of Contents for MGP 15 Dicom Theater
Page 1: ...Hardware Software Installation User Manual MGP 15 Dicom Theater...
Page 4: ...4 MGP 15 Dicom Theater This page intentionally left blank...
Page 9: ...9 Introduction...
Page 10: ...Introduction 10 This page intentionally left blank...
Page 23: ...23 Display Controller Installation...
Page 24: ...Display Controller Installation 24 This page intentionally left blank...
Page 27: ...BarcoMed 2MP2CP Overview 27 1920x1200 24 bit true color...
Page 32: ...Display controller installation 32 This page intentionally left blank...
Page 33: ...33 Display Controller Software Installation...
Page 34: ...Display Controller Software Installation 34 This page intentionally left blank...
Page 45: ...45 Display Controller Tools...
Page 46: ...Display Controller Tools 46 This page intentionally left blank...
Page 49: ...Barco Display Tab 49 Figure 20...
Page 50: ...Barco Display Tab CORONIS 3MP This page intentionally left blank 50...
Page 54: ...BarcoMed Driver Tab CORONIS 3MP This page intentionally left blank 54...
Page 62: ...BarcoMed Hardware Tab 62...
Page 63: ...63 Display Tools...
Page 64: ...Display Tools 64 This page intentionally left blank...
Page 74: ...Barco DPMS Screen Saver 74 This page intentionally left blank...
Page 75: ...75 MediCal Pro and Dicom Theater software...
Page 76: ...MediCal Pro and Dicom Theater software 76 This page intentionally left blank...
Page 92: ...Troubleshooting the stabilization process 92 This page intentionally left blank...
Page 93: ...93 Troubleshooting...
Page 94: ...Troubleshooting 94 This page intentionally left blank...