Troubleshooting the stabilization process
91
Troubleshooting the stabilization process
The projector stabilization has stopped unexpectedly
Possible causes:
•
No communication with projector. The projector may be
switched off, removed, or the RS-232 cable may be
disconnected at the projector’s or PC’s side.
•
No communication with optical sensor. The sensor
connection to the PC may be disconnected.
•
The optical sensor is not or no longer positioned on the
sensor patch.
•
The user has selected another signal input or another preset
for the projector.
In all of these cases, you must restore any problems in the
connections (if appropriate). Then you must run MediCal Pro and
select the proper Preset. This should restart the stabilization.
The PC does not boot or does not recognise the keyboard
after booting.
When a USB keyboard and the optical sensor are both connected
to the PC’s USB bus, there may be a problem at start-up. This is
not the case on all PCs.
A workaround is to connect the keyboard to the PS/2 connector.
Summary of Contents for MGP 15 Dicom Theater
Page 1: ...Hardware Software Installation User Manual MGP 15 Dicom Theater...
Page 4: ...4 MGP 15 Dicom Theater This page intentionally left blank...
Page 9: ...9 Introduction...
Page 10: ...Introduction 10 This page intentionally left blank...
Page 23: ...23 Display Controller Installation...
Page 24: ...Display Controller Installation 24 This page intentionally left blank...
Page 27: ...BarcoMed 2MP2CP Overview 27 1920x1200 24 bit true color...
Page 32: ...Display controller installation 32 This page intentionally left blank...
Page 33: ...33 Display Controller Software Installation...
Page 34: ...Display Controller Software Installation 34 This page intentionally left blank...
Page 45: ...45 Display Controller Tools...
Page 46: ...Display Controller Tools 46 This page intentionally left blank...
Page 49: ...Barco Display Tab 49 Figure 20...
Page 50: ...Barco Display Tab CORONIS 3MP This page intentionally left blank 50...
Page 54: ...BarcoMed Driver Tab CORONIS 3MP This page intentionally left blank 54...
Page 62: ...BarcoMed Hardware Tab 62...
Page 63: ...63 Display Tools...
Page 64: ...Display Tools 64 This page intentionally left blank...
Page 74: ...Barco DPMS Screen Saver 74 This page intentionally left blank...
Page 75: ...75 MediCal Pro and Dicom Theater software...
Page 76: ...MediCal Pro and Dicom Theater software 76 This page intentionally left blank...
Page 92: ...Troubleshooting the stabilization process 92 This page intentionally left blank...
Page 93: ...93 Troubleshooting...
Page 94: ...Troubleshooting 94 This page intentionally left blank...