1.9.1c Selecting mass flow units
1. Scroll to the desired
mass flow
units for the flow rate
display and press [enter]. The available units for this
prompt are determined by the selection made at
system units (see
Table 6
below).
Table 6: Available mass flow units
English
Metric
LB = pounds
KG = kilograms
KLB = thousands of LB
Tonne = metric tons (1,000 KG)
MMLB = millions of LB
Tons (2,000 LB)
2. Scroll to the desired
mass flow time
units for the mass
flow rate display and press [enter].
3. Scroll to the desired number of
Mdot decimal digits
(digits to the right of the decimal point in the mass flow
rate display) and press [enter].
4. Scroll to the desired
mass (totalizer)
units for the totalized
mass flow rate display and press [enter]. The available
units for this prompt are determined by the selection
made at the system units prompt.
5. Scroll to the desired number of
mass decimal digits
(digits to the right of the decimal point in the totalized
mass flow rate display) and press [enter].
Procedure options
After completing the previous steps, the XGF868i returns to
the global
program
window. Do one of the following:
• To continue programming, refer to
, to navigate to the desired menu.
• To leave the
keypad program
, press [escape] twice.
1.9.2 Setting up inputs and outputs
Set up the XGF868i inputs and outputs via the I/O submenu.
While following the programming instructions, refer to
and
The following
three submenus are included in this section:
•
Error
- program the meter’s response during an error
condition (see below)
•
Optn
- set up any option cards and the slot 0 analog
Setting up slot 0 and slot 1 inputs/outputs on
)
•
LCD
- set up the LCD display. Refer to
NOTE:
In this section, slot 1 appears as an option only if a
suitable option card is installed in slot 1.
Proceed to the appropriate section to program the
option selection made at the above prompt. Remember
to record all programmed data in
1.9.2a Selecting error handling
This menu option permits programming of the manner
in which the XGF868i outputs respond during an error
condition. See
Chapter 2, error codes
, in the
Service
manual
for a discussion of the built-in error codes. To
access this submenu:
NOTE:
The 2-channel meters have an additional option
for error handling.
1. In the
keypad program
, scroll to
prog
and press [enter].
2. In the
prog
menu, scroll to
globl
and press [enter].
3. In the global
program
menu, scroll to I/O and
press [enter].
4. Scroll to
error
and press [enter].
5. Scroll to the desired option for error handling and press
[enter]. If you select error level in mA, proceed to step
6 on the next page. (See
for a
description of error handling options available and how
the totalizers and analog outputs respond to them.)
NOTE:
The error responses listed in
apply only to the channelx measurement mode and to
the sum/difference measurement modes, if either or
both channels are in error. Error handling in the Average
measurement mode is set in
13
Summary of Contents for DigitalFlow XGF868i
Page 1: ...DigitalFlow XGF868i Panametrics flare gas flow transmitter Programming manual...
Page 2: ...ii...
Page 4: ...no content intended for this page iv...
Page 10: ...x...
Page 38: ...28...
Page 46: ...no content intended for this page 36...
Page 54: ...44...
Page 60: ...50...
Page 88: ...78...
Page 94: ...84...