1.8 Entering setup data
The signal limits, response times, mass flow and multiple
K factors for the XGF868i are specified via the setup
submenu. The following four submenus are included
in this section:
•
Signal
- set the parameters related to the transducer
signal (see below).
•
Default setup
- reset all parameters to default
values (
Initializing setup parameters - default setup
).
•
V averaging
- specify the response of the meter to
step changes (
Setting response time - V averaging
).
•
Advanced features
- enable mass flow and enter
K factors (
Using advanced features on page 11
).
While following the programming instructions, refer
. Record all programmed data in
Appendix B,
data records
.
1.8.1 Setting transducer signal parameters
Use this option to set the limits for the incoming signal
and other parameters affecting the transducer signal. For
example, the programmed signal strength low limit may be
used to determine the trigger point for an alarm.
CAUTION!
The signal default settings are suitable for
most applications. Consult with Panametrics
before changing any of these parameters.
Only after consulting the factory, complete the following
steps to program the signal parameters:
1. In the
keypad program
, scroll to
prog
and press [enter].
2. In the
prog
menu, scroll to CH1 or CH2 and press [enter].
3. In the channel
program
menu, scroll to setup and press
[enter].
4. Scroll to
signl
and press [enter].
5. Use
to choose the desired values.
Use the arrow keys to enter a value (or scroll to a value for
a parameter with menu options) and press [enter].
Procedure options
After completing the above steps, the XGF868i returns to the
channel SET UP window. Do one of the following:
• To continue programming, refer to
, to navigate to the desired menu.
• To leave the
keypad program
, press [escape] three times.
Table 4: Transducer signal settings
Transducer signal
parameters
Range
Default
value
Description
Signal low limit
-20 to 100
20
The E1:LOW SIGNAL error message appears when the signal strength falls
below the programmed SIGNAL LOW LIMIT value. See Chapter 2 in the
service manual for a discussion of error codes.
Correlation peak
limit
0 to 500
100
The E4: SIGNAL QUALITY error message appears when the signal quality
falls below the programmed COR. PEAK LIMIT value. See Chapter 2 in the
service manual for a discussion of error codes
Sou-
limit
1 to 50%
20%
The E2:SOUNDSPEED error message appears when the calculated
fluid sound speed differs from the fluid sound speed entered in the
Channelx-system menu by more than the programmed SOUNDSPEED
+- LIMIT value. See Chapter 2 in the service manual for a discussion of
error codes.
Velocity low limit
-500 to 500 ft/
sec (-150 to 150
m/sec)
-150 ft/sec
(-46 m/
sec)
The E3: VELOCITY RANGE error messages appears when the calculated
fluid velocity is less than the programmed VELOCITY LOW LIMIT value.
See Chapter 2 in the service manual for a discussion of error codes.
Velocity high limit
-500 to 500 ft/
sec (-150 to 150
m/sec)
150 ft/sec
(46 m/sec)
The E3: VELOCITY RANGE error messages appears when the calculated
fluid velocity exceeds the programmed VELOCITY HIGH LIMIT value.
See Chapter 2 in the service manual for a discussion of error codes.
8
Summary of Contents for DigitalFlow XGF868i
Page 1: ...DigitalFlow XGF868i Panametrics flare gas flow transmitter Programming manual...
Page 2: ...ii...
Page 4: ...no content intended for this page iv...
Page 10: ...x...
Page 38: ...28...
Page 46: ...no content intended for this page 36...
Page 54: ...44...
Page 60: ...50...
Page 88: ...78...
Page 94: ...84...