C.5.4 Entering input/output parameters
Enter the zero cutoff value and set up the temperature,
pressure and quality inputs via the input/output submenu.
While programming these parameters, refer to
of
Appendix D, PanaView menu maps
.
IMPORTANT
If an option card in slot 1 fails to appear in this menu, it
may be turned OFF. See
Setting up inputs and outputs on
page 65 (global-I/O-options
) for setup instructions.
C.5.4a Zero cutoff value
Near a zero flow rate, the XGF868i readings may fluctuate
due to small offsets caused by thermal drift or similar
factors. To force a zero display reading when there is
minimal flow, enter a
zero cutoff value
as described in
the following steps:
1. From the channel menu, highlight and double-click
on the input/output option in the center pane.
2. The program asks for the
zero cutoff
. Enter a value
from 0 to 1 ft/sec (0 to 0.30 m/sec) for the zero cutoff
and click [next item]. The recommended setting is
0.1 ft/sec (0.03 m/sec).
C.5.4b Temperature input
The XGF868i can use either a fixed temperature value
or a live temperature input to calculate the density for
the mass flow rate display.
1. Double-click on a
fixed
temperature value or to
set up the option card in
slot 1
that will supply the
live temperature input and press [next item].
NOTE:
If slot 1 contains an activated option card with an
analog input assigned to
temperature
or an
RTD
input,
slot 1 appears as an option at the above prompt. If the
process temperature is stable, a fixed value may be used,
but most applications require a live temperature input. If
there is no active option card for temperature, the meter
assumes you are using a fixed temperature.
2. Proceed to one of the following sections:
• If you selected
fixed
- proceed to step 3.
• If you selected
slot 1
- proceed to step 4.
3. Enter the known
fixed temp.
(process temperature)
and click [next item]. The meter will accept values
from –328 to 1832°F (–200 to 1000oC). Proceed to
base temperature
on the next page.
4. Select
input A
or
input B
and click [next item]. The inputs
were labeled during setup.
NOTE:
The set up of input A is used as an example.
Identical procedures would be used to set up input B.
C.5.4c Base temperature
1. Enter the
base temperature
and click [next item].
The ratio of this value to the actual temperature is
used to calculate the standard mass flow rate.
2. Do one of the following:
• If you selected pressure as the input type,
proceed to
pressure input
below.
• If you selected temperature as the input type,
proceed to base pressure below.
C.5.4d Pressure input
1. Double-click on a
fixed
pressure value or to set up
the option card in
slot 1
that will supply the live
pressure input.
NOTE:
If slot 1 contains an activated option card with
an input assigned to
pressure, slot 1
appears as an
option at the above prompt. If the process pressure is
stable, a fixed value may be used, but most applications
require a live pressure input. If there is no active option
card for pressure, the meter assumes you are using a
fixed pressure.
2. Proceed to one of the following steps:
• If you selected
fixed
- proceed to step 3.
• If you selected
slot 1
- proceed to step 4.
3. Enter the known
fixed
process
pressure
and click
[next item]. The meter will only accept values from
0 to 5000 psia. Proceed to
base pressure
below.
4. Double-click on
input A
or
input B
. The inputs were
labeled during setup.
NOTE:
The set up of input A is used as an example.
Identical procedures would be used to set up input B.
C.5.4e Base pressure
1. Enter the
base pressure
and click [next item]. The
ratio of this value to the actual pressure is used to
calculate the standard mass flow rate.
C.5.4f Low pressure switch
1. Click on
yes
or
no
to activate or deactivate the
low
pressure switch
software function and click [next item].
• If you selected yes, enter the
pressure limit
, the
low pressure switch set point, and click [next item].
The acceptable range is 0 to 5000 psia. The meter
will stop taking readings if the pressure drops below
this value.
PanaView returns to the channel
program
menu.
You have completed entering input/output parameters.
60
Summary of Contents for DigitalFlow XGF868i
Page 1: ...DigitalFlow XGF868i Panametrics flare gas flow transmitter Programming manual...
Page 2: ...ii...
Page 4: ...no content intended for this page iv...
Page 10: ...x...
Page 38: ...28...
Page 46: ...no content intended for this page 36...
Page 54: ...44...
Page 60: ...50...
Page 88: ...78...
Page 94: ...84...