Operating Manual Xenotest
®
220/220+
- 60 -
11
Technical Data
11. Technical Data
Incoming power:
Incoming power connection:
(1P/N/PE) AC
(2P/PE) AC
CEE (32 A, 3pole, 6 h)
Nominal voltage/frequency:
230 V ± 10%
Nominal power:
max. 5.1 kVA
Nominal power of the xenon lamp:
max. 2200 VA
Power of the xenon lamp:
max. 16 A
Dimensions and weight:
Dimensions in cm:
90 × 78 × 180
(width × depth × height)
Weight
ca. 280 kg
Light source:
Light source:
1 xenon lamp, air-cooled
Guaranteed life:
1500 hours
(1500 h at 65W/m
2
@ 300 – 400 nm)
Sample area
with XENOSENSIV
2310 cm²
Cooling air requirement:
Lamp unit
max. 150 m
3
/h
Test chamber:
max. 100 m
3
/h
Ambient conditions:
Room temperature:
max. 18 °C – 25 °C
Relative humidity:
50% (± 10%)
Water consumption:
Humidity control:
0.1 – 0.15 l/h
(approx. consumption under ISO 105-B02, normal conditions)
Noises:
Noise level:
<70 db (A)
Tab. 6
Tab. 7
Input limits
Input limits irradiance
Min
Max
CHT
20 °C
50 °C
RH
20 %
95 %
BST
20 °C
100 °C
BPT
20 °C
100 °C
300 – 400 nm
420 nm
Filter
Min Max
Min
Max
TM16
30
50 W/m
2
0.7
1.2 W/m
2
XENOCHROME 320 30
50 W/m
2
-
-
B04
30
50 W/m
2
-
-
The lamp stays off when entering „0“ (dark phase)