Operating Manual Xenotest
®
220/220+
- 42 -
6
Start up
6.10 Incoming power connection
WARNING – electrical shock!
Touching live parts can lead to life threaten-
ing electrical shock. Check plugs and power
cables for damage before connecting to the
incoming power. Damaged parts should not
be used for electrical connection!
The Xenotest 220/220+ is connected to an incoming volt
-
age of 230 V ± 10% (see technical data). The power supply
is made through a grounding plug (1P/N/PE), CEE (32 A,
3 pole, 6h). The electrical supply must be fused according
to regulations.
Connecting and turning on the instrument (fig. 35)
1. Before connecting to the incoming power, check wheth
-
er the values of the power supply network in the room
of operation matches the data on the rating plate on
the back of the instrument.
2. If the data for voltage (V) and maximum current (A) do
not match, the instrument should not be connected.
3. Insert the PE plug
2
into a properly grounded and
fused socket.
4. Make sure that no stress or strain is put on the power
cables.
5. Set the main switch
1
to the “
I
” position.
1
2
ATLAS
ESCAPE
ENTER
2
Start
Stop
3
1
5
6
4
8
9
7
0
Fig. 35