3.4 Water supply
The Xenotest 220/220+ requires purified water for hu
-
midifying the test chamber. Purified water circulates in a
closed pipe system which is fed from the internal 60-liter
water tank. The water tank can either be refilled manually
or connected to a water treatment system.
• Connection:
G 3/8 “ (inch)
The purified water must have the following quality
characteristics:
• Conductance:
< 5 µS/cm
• pH value:
6 – 8
• Max. silicate content:
0.2 ppm
NOTE – water quality:
Poor water quality not only leads to poor test results but
also affects the function and life of the components in
the instrument.
3
Requirements for the installation site
3.5 Transport
NOTE – Transport
The instrument may only be transported by experienced
service personnel using suitable equipment. The instrument
may not be lifted by the front or side doors for transport.
NOTE – lifting points (fig. 1)
The instrument may only be lifted at the lifting points
indicated in
fig. 1
.
Fig. 1
ATLAS
ESCAPE
ENTER
2
Start
Stop
3
1
5
6
4
8
9
7