Operating Manual Xenotest
®
220/220+
- 14 -
3
Requirements for the installation site
3.6 Space requirements
Position the instrument on a sufficiently stable, non-
flammable floor and align horizontally.
Weight of the instrument:
• Xenotest 220/220+
approx. 280 kg
NOTE – load at the installation site
If several instruments are to be installed in a room, ensure
the static load-bearing capacity of the floor.
An access area of at least 1 m is needed around the
machine to allow for operation, maintenance and repair
work, see
fig. 2
.
• Instrument dimensions:
90 cm × 78 cm × 180 cm (W × D × H)
CAUTION – overheating of the instrument!
The air exchange of the instrument takes place via air
inlet and outlet openings on the top of the instrument
and the side air vents. If the air exchange is blocked,
tests will be interrupted because the safety devices
will shut the instrument off. Also, repeatedly turning
the instrument on/off, will shorten the life of instrument
components. Make sure that the ventilation openings
are always clear!
• The room of installation must have a minimum height
of 2.5 m
• Comply with the minimum distances at the side and
back of the instrument. The safety and maintenance
distances can be found in
fig. 2