Operating Manual Xenotest
®
220/220+
- 30 -
6.1 Initial start up
NOTE – first start up
The instrument may only be started for the first time by an
Atlas technical service or authorized service representa-
tive. All the functioning components of the instrument are
checked during first-time start up.
6
Start up
6.2 Checking the instrument components
WARNING – electrical shock!
Touching live parts can lead to life threatening electri-
cal shock. Disconnect the instrument from the power
supply before checking components!
• Turn the instrument off at the ON/OFF switch
• Pull the electrical plug out of the socket and secure
against reconnecting
• Check whether the instrument is voltage free
Only instrument parts which are in optimal working con
-
dition ensure the functional reliability of the Xenotest
220/220+ instrument. Check the instrument components
and operational functions listed below for possible dam-
age and operability issues before starting or before every
new test run. Do not use damaged or worn components.
Checklist:
The following functions must be checked before starting:
• Correct combination of light filters
CAUTION – overheating!
The lamp filters assist in
reducing heat radiation. If
tests are performed without filters, the instrument can
overheat and the samples will be damaged. Only ignite
the xenon lamp when filters are properly installed.
• Number of operating hours of the xenon lamp
• Condition of the air filters for lamp cooling and test
chamber cooling
• Fit and tightness of the water hoses
• Condition of the water filter
• Fill level of the instrument’s internal water tank and
supply tank for unpurified water
• Condition of plugs and power cables