Troubleshooting
8
STERRAD VELOCITY Reader User’s Guide
49
Problem/Error Message
Possible Cause
Suggested Solution
No color change in the
chemical indicator for a
test biological indicator
may indicate that there’s
been an error in handling
or processing the biological
indicator. This may lead to
an incorrect result.
Improper handling
or processing.
Confirm that the user did
not inadvertently process a
control BI as a test BI. Process
a subsequent load with
another BI.
The biological indicator
has been removed from
well No. X. The incubation
period is void and has
been canceled.
BI removed before
reading process is
complete.
If a BI is removed before
the reading is complete the
reading stops, the results
are invalid, and a new test
will need to be conducted. Do
not touch or move BIs during
incubation. Remove BIs from
wells and perform proximity
calibration. Move reader away
from vibration sources. Ensure
BI was properly activated.
Our records indicate
that it’s been more than
two hours since sterilizer
cycle completion for this
biological indicator. This
may lead to an incorrect
result.
Power failure, loss,
delay.
Per BI IFU, incubate BI within
2 hours of sterilization to avoid
inaccurate results. Ensure
that the time is synchronized
between the reader and
the Sterilizer. BI is invalid,
reprocess with a new BI.
Our records indicate that
the scanned biological
indicator has already
been incubated. It cannot
be incubated again.
Scan barcode
from a previously
incubated BI.
Only incubate BIs which have
not been previously incubated.
Note: If the STERRAD VELOCITY
incubation cycle is canceled
after proceeding past the
BI activation screen, the
reader will consider the BI as
incubated and cannot incubate
that BI again.
Summary of Contents for Sterrad Velocity Reader
Page 1: ...User s Guide AB11321031 AB11321031 50 January 2021 ASP 2021 All rights reserved Reader...
Page 6: ...vi STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 14: ...8 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 36: ...30 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...34 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 52: ...46 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 66: ...THIS PAGE INTENTIONALLY LEFT BLANK 60 STERRAD VELOCITY Reader User s Guide...
Page 68: ...THIS PAGE INTENTIONALLY LEFT BLANK 62 STERRAD VELOCITY Reader User s Guide...
Page 70: ...THIS PAGE INTENTIONALLY LEFT BLANK 64 STERRAD VELOCITY Reader User s Guide...
Page 72: ...THIS PAGE INTENTIONALLY LEFT BLANK 66 STERRAD VELOCITY Reader User s Guide...