Operation
4
STERRAD VELOCITY Reader User’s Guide
19
Adding a Test BI
A test BI is a BI which has been exposed to a STERRAD Sterilization Cycle. The
accuracy of the BI result has been verified up to two hours post sterilization processing.
The test BI must be activated and processed in the reader within two hours of
completing the sterilization cycle.
9
Note:
As a precaution, when handling biological indicators that have been exposed
to hydrogen peroxide, please wear the appropriate PPE (chemical resistant latex,
PVC/vinyl, or nitrile gloves). Refer to the glove manufacturer’s instructions for use
for more information.
Touch an available well number or place the BI in an open well. The selected well
will then be used for reading. These steps can be repeated for any open well. Touch
Test
to add a test BI to the well. Scan the barcode on the BI cap label with the configured
barcode scanner. The reader is also able to use a handheld barcode scanner attached
to a USB port.
Figure 10.
Select a Well
Summary of Contents for Sterrad Velocity Reader
Page 1: ...User s Guide AB11321031 AB11321031 50 January 2021 ASP 2021 All rights reserved Reader...
Page 6: ...vi STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 14: ...8 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 36: ...30 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...34 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 52: ...46 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 66: ...THIS PAGE INTENTIONALLY LEFT BLANK 60 STERRAD VELOCITY Reader User s Guide...
Page 68: ...THIS PAGE INTENTIONALLY LEFT BLANK 62 STERRAD VELOCITY Reader User s Guide...
Page 70: ...THIS PAGE INTENTIONALLY LEFT BLANK 64 STERRAD VELOCITY Reader User s Guide...
Page 72: ...THIS PAGE INTENTIONALLY LEFT BLANK 66 STERRAD VELOCITY Reader User s Guide...