4
Operation
16
STERRAD VELOCITY Reader User’s Guide
Figure 5.
Select a Well
9
Note:
A new positive control BI should be performed at least once every 24 hours
using a BI from the same lot as the test BI.
Scan the barcode on the BI cap label with the scanner on the front of the reader.
The Reader is also able to use a handheld barcode scanner attached to a USB port. Hold the
biological indicator 10 cm (4 in) from the scanner with the barcode facing the scanner. Align
the barcode on the cap label with the center of the red light produced by the barcode
scanner. Slowly rotate the biological indicator back and forth until an audible beep is
heard. After successfully scanning the BI, enter the operator’s name and touch
Next
.
Figure 6.
Scan the BI Barcode
Summary of Contents for Sterrad Velocity Reader
Page 1: ...User s Guide AB11321031 AB11321031 50 January 2021 ASP 2021 All rights reserved Reader...
Page 6: ...vi STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 14: ...8 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 36: ...30 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...34 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 52: ...46 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 66: ...THIS PAGE INTENTIONALLY LEFT BLANK 60 STERRAD VELOCITY Reader User s Guide...
Page 68: ...THIS PAGE INTENTIONALLY LEFT BLANK 62 STERRAD VELOCITY Reader User s Guide...
Page 70: ...THIS PAGE INTENTIONALLY LEFT BLANK 64 STERRAD VELOCITY Reader User s Guide...
Page 72: ...THIS PAGE INTENTIONALLY LEFT BLANK 66 STERRAD VELOCITY Reader User s Guide...