4
Operation
20
STERRAD VELOCITY Reader User’s Guide
After successful scanning, enter the operator’s name and touch
Next
.
Figure 11.
Enter the Operator’s Name
Touch
Yellow
if the chemical indicator on the BI cap has changed color, and
then touch
Next
(refer to the “Chemical Indicator Result” section (Chapter 4) for a more
detailed explanation). If the color has not changed touch
Red / Pink
and refer to the
“Troubleshooting” section of this guide (Chapter 8).
Figure 12.
CI Color Change for a Test BI
9
Note:
If activating BI processed in the STERRAD 100S or without ASP ACCESS
continue to “Manual Entry of Cycle Information” (Chapter 4). Otherwise skip “Manual
Entry of Cycle Information” and proceed to “Activate the Test BI” (Chapter 4).
Summary of Contents for Sterrad Velocity Reader
Page 1: ...User s Guide AB11321031 AB11321031 50 January 2021 ASP 2021 All rights reserved Reader...
Page 6: ...vi STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 14: ...8 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 36: ...30 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 40: ...34 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 52: ...46 STERRAD VELOCITY Reader User s Guide THIS PAGE INTENTIONALLY LEFT BLANK...
Page 66: ...THIS PAGE INTENTIONALLY LEFT BLANK 60 STERRAD VELOCITY Reader User s Guide...
Page 68: ...THIS PAGE INTENTIONALLY LEFT BLANK 62 STERRAD VELOCITY Reader User s Guide...
Page 70: ...THIS PAGE INTENTIONALLY LEFT BLANK 64 STERRAD VELOCITY Reader User s Guide...
Page 72: ...THIS PAGE INTENTIONALLY LEFT BLANK 66 STERRAD VELOCITY Reader User s Guide...