5
Troubleshooting
54
STERRAD
®
100NX
®
User’s Guide
Call Your ASP Representative
If you encounter a problem or a system message that is not covered in the
this user’s guide, do not attempt to perform repairs or adjustments to the
STERRAD
®
100NX
®
Sterilizer.. Call your local ASP Representative for
more information.
Unable to Evacuate Chamber . . .
Ensure Load is Dry before Starting
Cycle
Remove the load and run diagnostics. Reboot the system.
If the problem persists, call your ASP Representative.
Unable to Open One Second Data File
Replace the PCMCIA card with a new one.
UV Path Is Blocked, Open Door And
Clear Pathway
Object is blocking the UV path. Verify that the shelves and/or
the load are not blocking the path. If the problem persists, call
your ASP Representative.
Warming Up, Please Wait…
No action needed.
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear chemical resistant latex, PVC (vinyl), or nitrile gloves whenever handling a
load after a cycle cancellation. Hydrogen peroxide liquid may be present on the
load or in the chamber.
Would You Like to Troubleshoot?
Remove the load and run diagnostics. If the problem persists,
call your ASP Representative.
If this message appears. . .
Do this . . .
DRAFT
Summary of Contents for Sterrad 100NX
Page 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Page 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Page 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Page 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Page 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Page 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Page 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Page 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Page 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...