A
Specifications
104
STERRAD
®
100NX
®
User’s Guide
Humidity
Operating: 10% – 85% up to 30
o
C.
linearly decreasing from 85% at 30
o
C to 70% at 40
o
C non-condensing
Storage: 10% – 100% (rainfall will be permitted).
Altitude/Pressure
Operating altitude up to 3095 m (10,152 ft.).
Atmospheric pressure 520-775 torr
Cycle temperature
47°C – 56°C (116.6°F – 132.8°F)
Cycle time
STANDARD Cycle - 47 minutes
DUO Cycle - 60 minutes
EXPRESS Cycle - 24 minutes
FLEX Cycle - 42 minutes
Connectors
Network: RJ45; Barcode reader: USB.
Printer: USB
Main Power Cable
12 AWG (4 mm
2
), 3 m (9.84 feet) long, 5 conductors
NEMA L21 - 30P (USA and Canada.)
Each country is responsible for installing an appropriate 3 phase plug according to
their facility’s power requirements. The plug must match the phase rotation.
Mobility
On wheels
Venting Requirements
None required
Gas Tank
Requirements
None required
Heat Generation
Idle: 289 BTU/hour
In operation: 8,189 BTU/hour maximum
RF Generation
Portable and mobile RF communications equipment can affect medical
Electrical Equipment.
DRAFT
Summary of Contents for Sterrad 100NX
Page 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Page 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Page 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Page 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Page 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Page 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Page 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Page 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Page 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...