3
Load Preparation
24
STERRAD
®
100NX
®
User’s Guide
Devices with internal parts, such as sealed bearings, that cannot be
immersed, may present difficulties in cleaning and should not be
processed in the STERRAD
®
100NX
®
Sterilizer.
Implants for which the manufacturer has not specifically recommended
sterilization in the STERRAD
®
100NX
®
Sterilizer.
Do Not Process in the EXPRESS Cycle
Items made of nylon cannot be processed in the EXPRESS Cycle.
Items made of Kraton
®
cannot be processed in the EXPRESS Cycle.
Items made of Polyurethane cannot be processed in the EXPRESS Cycle.
Items with mated Delrin
®
surfaces cannot be processed in the EXPRESS
Cycle.
Items with mated anodized aluminum surfaces cannot be processed in the
EXPRESS Cycle.
Items with mated Radel
®
surfaces cannot be processed in the EXPRESS
Cycle.
Items with mated Ultem
®
surfaces cannot be processed in the EXPRESS
Cycle.
Items with lumens cannot be processed in the EXPRESS Cycle.
Do Not Process in the DUO Cycle
Items with mated anodized aluminum surfaces cannot be processed in the
DUO Cycle.
DRAFT
Summary of Contents for Sterrad 100NX
Page 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Page 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Page 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Page 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Page 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Page 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Page 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Page 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Page 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...