3
Load Preparation
28
STERRAD
®
100NX
®
User’s Guide
Do not use paper pouches or sterilization wraps containing cellulose
or cotton.
Do not use any wraps or packaging that are not approved by ASP or
materials listed in the “Items Not To Be Processed” section. In the USA,
use only FDA-cleared polypropylene wraps.
Properly arrange the items or the scope in a tray to ensure adequate
diffusion of hydrogen peroxide throughout the load.
Place peel pouches on edge, if possible. Arrange them so that the
transparent side of a pouch faces the opaque side of the next pouch.
Do not stack pouches on top of each other.
Do not stack instruments inside the trays. Do not stack trays. Do not stack
trays within trays. Do not wrap instruments within a wrapped tray.
If you are using rigid containers cleared by the FDA for use in the
STERRAD
®
100NX
®
Sterilizer, follow the Instructions for Use provided
by the rigid container manufacturer. Verify that the rigid containers are
cleared for use in each sterilization cycle. Remember the following:
Do not stack instruments inside the containers.
Do not stack containers.
Do not stack containers within containers.
Do not wrap instruments within the containers.
Place STERRAD
®
Chemical Indicator Strips inside trays and pouches
as needed.
Loading
Do not allow any item to touch the walls of the sterilization chamber,
door, or electrode.
STANDARD and FLEX Cycles Loading Preparation
The STERRAD
®
100NX
®
Sterilizer
STANDARD
and
FLEX
Cycles
were validated using a load weight of 4.9 kg (10.7 lbs) per shelf. When
constructing your load, the total weight of the load to be sterilized should
not exceed 9.7 kg (21.4 lbs).
DUO Cycle Loading Preparation
The STERRAD
®
100NX
®
Sterilizer
DUO
Cycle was validated using a
total load weight of 6.0 kg (13.2 lbs). When constructing your load, the
total weight of the load to be sterilized should not exceed 6.0 kg (13.2
lbs).
DRAFT
Summary of Contents for Sterrad 100NX
Page 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Page 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Page 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Page 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Page 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Page 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Page 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Page 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Page 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...