7
Maintenance
70
STERRAD
®
100NX
®
User’s Guide
Cleaning the Hydrogen Peroxide Monitor
Detector Lens
The hydrogen peroxide monitor lens is located on the input side of the
sterilizer. The lens must be kept clean. Wipe off the lens once every three
months or when an accumulation of debris is noted. This is shown in the
following figure.
Figure 33. Cleaning the Hydrogen Peroxide Monitor Lens.
1. Always use a lint-free cloth to clean the lens.
2. Moisten the cloth with isopropyl alcohol. Never use an abrasive cleanser.
3. Wipe the lens to remove any accumulated debris.
DRAFT