Access Levels and Supervisor Tasks
9
STERRAD
®
100NX
®
User’s Guide
97
Diagnostic Tests
*
Times are approximate. If a failure is detected, the time may be extended.
The ten tests take approximately 23 minutes and 10 seconds to complete.
When the series of tests is complete, the sterilizer creates and stores a
diagnostics file and prints a report. When printing is complete, the
Additional Utilities menu is displayed.
Service Functions
The
Service Functions
button is reserved for use by ASP Service
Representatives.
Order
Test Name
What is tested
Average Time
to Run*
1
Power Supply Test
DC power supply
30 sec.
2
Vacuum Test
Vacuum pump, pressure sensors and valves
(vacuum control, vent, inlet, transition, delivery,
and oil return)
12 min.
3
Door Test
Door Motor and sensors
20 sec.
4
Plasma Test
Plasma electrical subsystem
Electrode integrity
3 min. 20 sec.
5
Cassette Test
Cassette electro-mechanical subsystem RFID
Reader
5 min.
6
H
2
O
2
Sensor Test
Ultraviolet lamp and detector
20 sec.
7
Display Test
Touch screen and display function
20 sec
8
Printer Test
Printer function
10 sec.
9
Fan Test
Fan function
10 sec.
10
Sound Test
Loudspeaker function and volume
1 min.
DRAFT
Summary of Contents for Sterrad 100NX
Page 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Page 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Page 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Page 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Page 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Page 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Page 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Page 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Page 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...