Troubleshooting
5
STERRAD
®
100NX
®
User’s Guide
51
Cassette System Timeout
When Piercing
The delivery subsystem is not responding. Remove the load and
run diagnostics.
Cassette Used, Disposing
The inserted cassette does not have any unused cells remaining.
Insert a new cassette.
Control Available On Other Side
Use the other side of the sterilizer to access the touch screen.
Cycle Canceled By Operator
The operator canceled the cycle. Repackage the load. Restart the
cycle after cancellation is complete.
Delivery System Not Ready
Delivery system not responding. Remove the load and
run diagnostics.
Disposing Cassette
No action required.
Door Opened
Door open sensor malfunction. Remove the load and
run diagnostics.
Door Sensor Fault
Door malfunction. Reboot the system. If the message persists,
call your ASP Representative.
Function Available on Other Side Only
Control of the sterilizer is on the other side of the unit.
H
2
O
2
Adjustment In Progress
The intensity of the UV lamp is being adjusted. Wait 5 minutes
for adjustment to be completed.
H
2
O
2
Bulb Warming Up, Please Wait…
No action needed.
H
2
O
2
Bulb/Detector Fault
H
2
O
2
detector malfunction. Run diagnostics.
H
2
O
2
Curve Area Too Low. Verify That
The Load Is Not Blocking The UV
Light Path
Load is absorbing too much peroxide. Remove absorbing
materials from the load, repackage, and restart the cycle. If the
problem persists, call your ASP Representative.
H
2
O
2
Monitor Failure
H
2
O
2
bulb or detector malfunction. Remove the load and
run diagnostics.
WARNING! HYDROGEN PEROXIDE MAY BE PRESENT.
Wear chemical resistant latex, PVC (vinyl), or nitrile gloves whenever handling
a load after a cycle cancellation. Hydrogen peroxide liquid may be present on
the load or in the chamber.
H
2
O
2
Peak Too Low
The load is absorbing too much peroxide. Remove the absorbent
materials from the load, repackage, and restart the cycle. If the
problem persists, call your ASP Representative.
H
2
O
2
Rate Constant Too High
The load is decomposing the H
2
O
2
. Check the load for absorbent
materials. Repackage, and restart the cycle. If the problem
persists, call your ASP Representative.
If this message appears. . .
Do this . . .
DRAFT
Summary of Contents for Sterrad 100NX
Page 1: ...STERRAD 100NX Sterilization System User s Guide Ref 99994 M 99994_03 May 2014 D R A F T...
Page 21: ...Load Preparation 3 STERRAD 100NX User s Guide 19 D R A F T...
Page 22: ...3Load Preparation 20 STERRAD 100NX User s Guide D R A F T...
Page 23: ...Load Preparation 3 STERRAD 100NX User s Guide 21 D R A F T...
Page 49: ...Operation 4 STERRAD 100NX User s Guide 47 Cycle Completion Flowchart D R A F T...
Page 76: ...7Maintenance 74 STERRAD 100NX User s Guide D R A F T...
Page 80: ...8Reports and Files 78 STERRAD 100NX User s Guide D R A F T...
Page 104: ...9Access Levels and Supervisor Tasks 102 STERRAD 100NX User s Guide D R A F T...
Page 108: ...A Specifications 106 STERRAD 100NX User s Guide D R A F T...