7. Assembly
7.1
Operating site
Please follow these instructions when determining the operating site:
›
The installation location of the device is variable. These can be:
1) Measuring point at a pressure line.
This measuring point should be representative of the machine. The device itself provides the necessary volume flow via an
integrated volumetric flow controller.
2) Measuring a sample from a tank or a bottle over the pump.
Connect the device to a pressure line in the off-line flow by means of a T-junction.
›
At the connection point, constant pressure conditions should prevail. The pressure may vary, but pressure peaks or strong
fluctuations must not occur.
›
The connection to the control line is recommended, alternatively the filter or cooling circuit is suitable.
7.2
Hydraulic connection
CAUTION
Danger due to improper handling
Material damage.
Always connect the oil return first.
The device is provided with three connections.
1
2
3
1 Oil outlet CPC-LC
2 Pressure line Minimess M16x2 (for operation without pump), max. 320 bar / 4600 psi
3 Suction line CPC-LC (for operation with internal pump)
Fig. 8: Hydraulic connections