Notes on measurement
These notes are stored in the device.
After a measurement the following information can be given:
Alarm condition RED
No. Description
0
Low oil level
4
Free water (>95%)
5
Extreme water content (>75%)
6
High temperature (>80ºC)
9
High volume flow index (<500 ml/min)
10
Low volume flow index (<50 ml/min)
48
Particle measurement laser current too high
49
Laser current too low
50
Photodiode voltage too low
51
Photodiode voltage too high
No alarm display / GREEN
No. Description
25
Temperature outside the measuring range
26
Humidity outside the measuring range
27
Conductivity outside the measuring range
28
Permittivity outside the measuring range
53
Incorrect temperature measurement
52
Incorrect humidity measurement
54
Incorrect conductivity measurement
55
Incorrect permittivity measurement
7
High average temperature
49
Defective condition sensor
Alarm condition YELLOW
No.
Description
20
High water content (>50%)
Please also refer to Chapter 16 - Troubleshooting
›
Measured values oil contamination
ISO 4, 6 ,14, 21 Cleanliness class according to ISO
SAE 4, 6, 14, 21 Cleanliness class according to SAE
NAS
Cleanliness class according NAS
GOST
Cleanliness class according GOST
Conc 4, 6, 14, 21 Concentration
F
Index
Flow index
›
Info text