5-6
Veriti™ Thermal Cycler User Guide
Chapter 5
Utilities
3. Download the new firmware to a USB drive:
a. Insert the USB drive into the USB port on the computer.
b. Click the link in the Software column to download the file. Select the USB
drive as the location for the saved file.
c. Remove the USB drive from the computer when the download is complete.
To upgrade the instrument with the new firmware:
“Backing Up Run Methods and Other Settings” on
to save your run methods and settings from the Veriti Thermal Cycler.
2. Insert the USB drive with the new firmware in the USB port on the front of the
Veriti Thermal Cycler,
3. Log in as a user with Administrator access.
4. In the Main Menu screen, touch
Settings Menu
.
5. In the Settings Menu screen, touch
Admin Menu
.
6. In the Admin Menu screen, touch
Upgrade Firmware
.
The Firmware Upgrade screen opens:
7. Touch
Source File
to display a list of the files on the USB drive. Veriti
™
system
firmware files end with “.update”.
8. Touch the name of the file with the new firmware, then touch
Select
.
9. Touch
Upgrade
, then
Yes
.
When the upgrade process is complete, a message box opens.
10. Restart the instrument.
Summary of Contents for Veriti
Page 1: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 2: ......
Page 3: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 8: ...vi Veriti Thermal Cycler User Guide ...
Page 12: ...x Veriti Thermal Cycler User Guide Preface ...
Page 78: ...3 26 Veriti Thermal Cycler User Guide Chapter 3 Performing a Run ...
Page 82: ...4 4 Veriti Thermal Cycler User Guide Chapter 4 Troubleshooting ...
Page 96: ...5 14 Veriti Thermal Cycler User Guide Chapter 5 Utilities ...
Page 106: ...A 4 Veriti Thermal Cycler User Guide Chapter A Instrument Specifications ...
Page 118: ...C 8 Veriti Thermal Cycler User Guide Chapter C Predefined Run Methods ...
Page 123: ......