3-24
Veriti™ Thermal Cycler User Guide
Chapter 3
Performing a Run
Using the Run
Monitor Button
The Run Monitor button (
) is displayed at the top of the Run Monitor
screen. If you leave the Run Monitor screen during a run, this button also appears at
the top of all menu screens.
Figure 3-1
The Run Monitor button
You can touch the Run Monitor button to return to the Run Monitor screen at any
time.
If an error occurs during the run, the area showing the current stage, in the middle of
the button, is replaced by an alert icon:
Figure 3-2
Alert icon, shown in the Run Monitor button when an error occurs
If the Power Fails
If a power failure occurs during a run, the instrument attempts to continue the run
when the power returns. The run begins at the first step of the stage where the power
failed.
For example, the power fails during step 2 of the tenth cycle for a stage with three
steps and 20 cycles. After the power returns, the run starts at step 1 of the tenth cycle
and continues for 11 more cycles.
Note:
If the time between the initial power failure and the time the power returns is
more than 18 hours, the run is not continued.
Summary of Contents for Veriti
Page 1: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 2: ......
Page 3: ...Applied Biosystems Veriti Thermal Cycler User Guide ...
Page 8: ...vi Veriti Thermal Cycler User Guide ...
Page 12: ...x Veriti Thermal Cycler User Guide Preface ...
Page 78: ...3 26 Veriti Thermal Cycler User Guide Chapter 3 Performing a Run ...
Page 82: ...4 4 Veriti Thermal Cycler User Guide Chapter 4 Troubleshooting ...
Page 96: ...5 14 Veriti Thermal Cycler User Guide Chapter 5 Utilities ...
Page 106: ...A 4 Veriti Thermal Cycler User Guide Chapter A Instrument Specifications ...
Page 118: ...C 8 Veriti Thermal Cycler User Guide Chapter C Predefined Run Methods ...
Page 123: ......