C-6
Veriti™ Thermal Cycler User Guide
Chapter C
Predefined Run Methods
LMS2 Run Method
Use this run method for ABI P
RISM
®
Linkage Mapping Set Version 2.5 (LMS2)
thermal cycling. Linkage Mapping Set Version 2 uses over 400 fluorescently labeled
PCR primer pairs for analysis of select microsatellite loci from the Généthon human
linkage map.
Time Release Run Method
Use this run method for use with AmpliTaq Gold
®
DNA Polymerase. This method
minimizes the pre-PCR activation step and adds a minimum of 10 additional cycles,
allowing for slow activation of the enzyme during cycling. This provides a simple
method where polymerase activity increases more slowly as product accumulates,
improving specificity.