The SimpliAmp
™
Thermal Cycler has a Pollution Degree rating of 2. It may be
installed in an environment that has non-conductive pollutants only, such as dust
particles or wood chips. Typical environments with a Pollution Degree II rating are
laboratories and sales and commercial areas.
The safety of SimpliAmp
™
Thermal Cycler use was tested for altitudes up to 6000 ft.
Technical specifications
The following table displays the system specifications for the SimpliAmp
™
Thermal
Cycler.
Table 9
SimpliAmp
™
Thermal Cycler technical specifications:
Feature
Specification
Maximum block ramp rate
[1]
4°C/sec
Maximum sample ramp rate
[1]
3°C/sec
Temperature accuracy
±0.25°C (35 to 99.9°C)
Temperature range for
protocol run
0°C to 100.0°C
Temperature non-uniformity
[2]
<0.5°C
Dimensions
• Height: 21 cm (8.27 in)
• Width: 24.0 cm (9.45 in)
• Depth: 46 cm (18.11 in)
Weight
8.5 kg (18.3 lb)
PCR volume range
• Supported: 10 - 100 µL
• Settable: 1 - 100 µL
Instrument memory
USB, on-board
Display interface
8" color TFT LCD
Power
100-240 V, 50-60 Hz Max: 600 W
VeriFlex
™
blocks
• 3 VeriFlex
™
zones
• Supported 10°C (5°C zone-to-zone)
• Settable
[3]
20°C (10°C zone-to-zone)
Ambient humidity
15% - 80% Relative Humidity, non-condensing
(acceptable range)
Ambient operating
temperature
15°C to 30°C (acceptable range)
Pollution
Altitude
Appendix D
Instrument specification and layout
Technical specifications
D
64
SimpliAmp
™
Thermal Cycler User Guide
Summary of Contents for SimpliAmp Thermal Cycler