Upgrade the firmware
1.
Open the About Instrument screen to determine the version number of the
existing firmware. To open the About Instrument screen, in the Home screen,
touch Settings. In the Settings screen, touch About Instrument.
2.
Record the version number of the firmware, then touch Done.
3.
On a computer connected to the Internet, go to thermofisher.com to determine
the version number of the new firmware.
4.
Click Product Support, then Technical Resources
u
Product Support
u
Software, Patches & Updates
5.
Select SimpliAmp
™
Thermal Cycler in the list, then click Updates & Patches.
6.
Find the appropriate file. If the version number is:
• The same as the current version (from step 3), you do not need to upgrade the
firmware.
• Different from the current version (from step 3), follow the steps below to
download and upgrade the firmware.
7.
Insert a USB drive into the USB port on the computer to download the new
firmware to the USB drive.
8.
Click the link in the Software column to download the file. Select the USB drive
as the location for the saved file.
9.
Remove the USB drive from the computer when the download is complete.
To upgrade the instrument with the new firmware:
1.
Insert the USB drive with the new firmware in the USB port.
2.
In the Home screen, touch Settings, or touch the shortcut to the Upgrade
Firmware screen if you have enabled the USB pop-up.
3.
In the Settings screen, touch Maintenance & Services.
Chapter 4
Maintain the instrument
As-needed maintenance
4
SimpliAmp
™
Thermal Cycler User Guide
35