3-9
A Look at the CYCL Program
In the following steps, you will look at the main displays in the
CYCL program. In each display, you will see the default
parameters (e.g., times and temperatures) that make up a
common PCR cycle. Using the
STEP
key, you will quickly step
through all the displays and accept the default values. Keep in
mind, however, that default values can be changed, which you
will do later on in this tutorial.
1
The first screen in the CYCL program should still be on your
display:
This display sets the number of setpoints or incubation
temperatures in one cycle. The cursor marks the default value, 3,
which specifies three-temperature PCR cycling.
2
Press
STEP
. This accepts the default value of three set-points and
moves you to the next display:
This display sets the ramp time, temperature, and hold time for
Setpoint 1 — typically used for template denaturation.
The temperature of the first PCR setpoint is 94 degrees C and is
held for 30 seconds.
3 Temperature PCR
Cursor
Setpt #1 Ramp 0:00
94.0C Hold
0:30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...