7-26
Changing the Temperature Interval for the Setpoint Clock
Whenever you run a program, there is an interval before the
setpoint temperature is attained. The clock starts counting the
hold time at the beginning of this interval.
For example, if the setpoint temperature is 95 degrees C, and the
temperature interval is 1 degree C, the clock will actually start
counting at 94 degrees C. Figure 7-7 illustrates this example.
Figure 7-7
Diagram of a 1-degree C temperature interval.
The default value for the temperature interval is 1 degree C. The
Configuration function lets you change this value.
CAUTION
Always check this parameter before you start a run.
To change the temperature interval for the setpoint clock:
1
Select CONFIG on the Utilities menu as described on page 7-3.
The following display appears:
•
Sample
Temperature
Time
•
95 C
o
94 C
o
Setpoint clock
starts here
1 C Interval
o
Setpoint Tem-
perature
94 C
o
95 C
o
C
o
Configuration
EDIT-PRINT
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...