2-21
Chiller Test
The chiller test verifies that the refrigeration system is operating
correctly.
CAUTION
Before you begin this test, make sure you place an
empty MicroAmp Tray (MicroAmp Tray/Retainer Set, Part No.
N801-0530) on the sample block, then slide the heated cover
forward and turn the cover knob clockwise until the white mark on
the knob lines up with the white mark on the cover.
1
The following display should appear. If not, follow Steps 1 through
3 in the heater test procedure on page 2-19.
2
To access the chiller test, press
3
and press
ENTER.
The following display appears and the test begins:
The test takes from 3 to 12 minutes to complete. If “Chiller Test
Passed” appears on the display as shown below, the refrigeration
system is operating correctly. If “Chiller Test Failed” appears on the
display, notify an Applied Biosystems Service Engineer.
3
Press
STOP
to exit the test. The next test (Test #4) will appear on
the display. Press
STOP
again to return to the Utilities menu.
Enter Diag test # 1
REVIEW HISTORY FILE
Enter Diag test # 3
CHILLER TEST
Chiller Test Passed
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...