6-2
Loading Samples
The GeneAmp PCR System 9600 uses a 96-position sample tray
(MicroAmp Tray) and retainer (MicroAmp Retainer) for 0.2-mL
MicroAmp Reaction Tubes.
Using the 96-Position Sample Tray with MicroAmp Reaction Tubes
To load the 96-position MicroAmp Tray, follow the steps below
(refer to Figure 6-1).
CAUTION
The GeneAmp PCR System 9600 is designed for
oil-free
operation. It does
not
use mineral oil or glycerine in the
sample block. The MicroAmp Reaction Tubes fit tightly in the wells
and a heated cover exerts an even pressure on all tubes.
No
mineral oil is used on the sample surface because the heated
cover eliminates condensation on the tubes.
1
Place the MicroAmp Base on a flat surface such as a table.
Do not
place the base in the sample block.
2
Place the MicroAmp Tray onto the base so that the well numbered
A1 is located at the upper left corner of the tray. The cutout in the
outside rim of the tray will then be located on the left side of the tray.
This orients the tray for the proper fit into the sample block.
3
Place MicroAmp Reaction Tubes in the tray. Note that the wells in
the tray are labeled left to right from 1 through 12 and A through H
from back to front.
Note
The above procedure is most useful if you are loading a dozen
or more samples. If you are only loading a few samples, use the sample
tray base as a working tray in which to load your sample tubes, then
transfer your tubes to a sample tray already loaded in the instrument.
Do not use the MicroAmp Retainer if you want to remove a single tube
without removing the caps from all the tubes. The caps will not fit
through the retainer.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...