2-20
3
Press the
OPTION
key twice to move the cursor under DIAG and
press
ENTER
.
A display referring to the first diagnostic test appears:
4
Change the "1" to a "2" by pressing
2.
Then press
ENTER.
The following display appears and the test begins:
If “Heater Test Passed” appears on the display as shown below,
proceed to the chiller test. If “Heater Test Failed” appears on the
display, notify an Applied Biosystems Service Engineer.
5
Press
STOP
to go to the next test (Test #3).
Next, you will run the chiller test, which is accessible from the first
diagnostic display.
Enter Diag test # 1
REVIEW HISTORY FILE
Enter Diag test # 2
HEATER TEST
Heater Test Passed
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...