5-35
You can accept any of the default values for Setpoint
#1 by pressing the
ENTER
key, or you can change
the values using the following steps:
1
For the ramp time, use the numeric keys to specify
the time (in minutes and seconds). For example, to
specify a ramp time of one minute, press
1 0 0
, then
press
ENTER
.
2
For the setpoint temperature, enter the temperature
in decimal form. You do not type a decimal point. For
example, to specify 96.0 degrees C, press
9 6 0
, then
press
ENTER
.
3
Specify the hold time in minutes and seconds, then
press
ENTER
. For example, to specify fifteen
seconds, press
1 5
, then press
ENTER
.
Setpoint #2 in a typical cycle program is used for
primer annealing. This display sets the ramp time,
target temperature, and hold time for the second
setpoint (Setpt #2). For a three-temperature PCR,
the default temperature is 55 degrees C. For a two-
temperature PCR, the default is 68 degrees C. The
other default values are shown in the display to the
left.
You can accept any of the default values for Setpoint
#2 by pressing the
ENTER
key, or you can change
the values using the same procedure as that for
Setpoint #1 (see above).
Setpt #2 Ramp 0:00
55.0C Hold
0:30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...