3-17
Defining Values for Setpoint #2
1
Use the default ramp time by pressing
ENTER
2
Enter 52.5 degrees C for the temperature by pressing
5
, then
2
,
then
5
. Then press
ENTER
.
3
We will not change the hold time, Press
STEP
to skip over this
entry,
The display for Setpoint #3 appears.
Defining Values for Setpoint #3
1
Use the default ramp time by pressing
ENTER
.
2
Enter 74.5 degrees C for the temperature. Press
ENTER
.
3
We will not change the hold time, Press
STEP
to skip over this entry
and move to the next display:
Setpt #2 Ramp 0:00
55.0C
Hold 0:30
Setpt #3 Ramp 0:00
72.0C
Hold 0:30
Total cycles = 25
Pause during run? NO
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...