3-14
Customizing the CYCL Program
In the steps below, you will customize the CYCL program by
changing some of the default parameters. Although we suggest
some new temperatures and times to use, you may want to
choose values relating to a current PCR application in your lab.
Defining the Number of Setpoints
1
The first display in the CYCL program (shown below) should appear
on your display. If it does not, press
STOP
one or two times to
return to the main menu. Then refer to “Getting to the CYCL
Program” on page 3-6.
2
In this display, you specify the number of setpoints to be used in the
program. (For example, if you wanted to perform two-temperature
PCR, you would press
2
.) For this tutorial, you will set up a three-
temperature PCR program, so just press
ENTER
to accept the de-
fault value of 3.
The display for Setpoint #1 appears.
3 Temperature PCR
Setpt #1 Ramp 0:00
94.0C Hold
0:30
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...