7-30
Setting Sample Tube Time Constants
There are two additional parameters in the Configuration function
that let you change the time constants for MicroAmp Reaction
Tubes and Thin Walled GeneAmp Reaction Tubes. We
recommend that you do not change the time constant values
unless instructed to do so by future documentation.
MICRO Time Constant
To change the sample tube time constant value for MicroAmp
Reaction Tubes:
1
Select CONFIG on the Utilities menu as described on page 7-3.
The following display appears:
2
Press
ENTER
, then press
STEP
until the following display appears:
3
The default condition of this parameter is no. Press the
OPTION
key. “NO” on the display will change to “DEFAULTS.”
4
Press the
OPTION
key again. “DEFAULTS” will change to
“MODIFY.” Pressing the
OPTION
key again will change “MODIFY”
back to “NO.”
Configuration
EDIT-PRINT
Set MICRO tube time
constants? NO
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for GeneAmp PCR System 9600
Page 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 35: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 43: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 79: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 107: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 167: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 251: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 275: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 281: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 294: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...