Appendix A: Technical Data
72
D52IB001EN-G
Sample Holder
CCU 100 Dimensions:
CCU 100 Electrical Data (For use with TTK 600):
Cooling water requirements:
Compressed-air requirements of air cooling:
Compressed-air requirements for LN2 cooling:
Sample holder surface:
Nickel
Sample cavity area (L x W):
14 x 10 mm
Sample cavity depth:
0,8 mm, 0,2 mm or flat
(other depths on request)
Zero background holder:
Si(510), thickness 0.5 mm
Width x Depth x Height:
450 x 410 x 180 mm
Weight:
approx. 13 kg
Voltage (mains):
AC 100-240 V
Frequency (mains):
50 ... 60 Hz
Power consumption:
max. 250 VA
Mains fuses:
2 x T 6,3A H 5x20 mm (ceramic tube)
Overvoltage category:
II according to EN 61010-1
Voltage output:
DC 48 V
Current output:
max. DC 4 A
Heater fuse:
2 x T4 A (SMD)
Flow rate:
0.7 to 1.5 l/min
Temperature:
15 to 25 °C
Pressure:
2 to 8 bar
Pressure:
6 - 8 bar rel.
Flow rate:
120 l/min at 6 bar
pressure dew point (@ 7 bar):
- for operating temperature
+5 °C
- for operating temperature < +5 °C
at 7 bar
< +25 °C
< +10 °C
Purity (acc. DIN 8573-1):
solid particle class 3 or better
oil class 2 or better
Pressure:
> 2.5 bar rel.
Flow rate:
70 l/min at 2.5 bar
Summary of Contents for TTK 600
Page 2: ......
Page 74: ...Appendix B Sample Chamber Dimensions 74 D52IB001EN G Appendix B Sample Chamber Dimensions ...
Page 75: ...Appendix C Declaration of Conformity D52IB001EN G 75 Appendix C Declaration of Conformity ...
Page 77: ...Appendix D Spare Parts and Accessories D52IB001EN G 77 176383 COOLING MODE SWITCHING VALVE ...